CLIA Waived Testing Daily Quality Control Log
Daily QC log for CLIA-waived point-of-care tests that records control results, lot numbers, expiration checks, and corrective actions in one place. Use it to document each testing session and catch invalid runs before patient results are reported.
Trusted by frontline teams 15 years of frontline software AI customization in seconds
Built for: Urgent Care · Primary Care Clinics · Retail Health Clinics · Ob/gyn Practices · Small Medical Labs
Overview
This CLIA Waived Testing Daily Quality Control Log is for documenting the control checks that must happen before waived point-of-care results are used. It captures the test name, operator, instrument or analyzer ID, reagent and control lot numbers, expiration dates, positive and negative control outcomes, and the final QC decision for that session.
Use it when your site performs waived testing that requires daily QC, lot-to-lot verification, or any manufacturer-directed control run. It is especially useful in clinics, urgent care centers, retail health sites, and small labs where multiple operators may use the same device and where traceability matters during inspection or internal review.
Do not use this log as a patient result worksheet or as a substitute for the manufacturer’s package insert. If the test system has a different QC frequency, different control levels, or special storage requirements, those instructions govern the workflow. It is also not the right tool for moderate or high complexity testing, instrument maintenance records, or competency documentation. The log is most valuable when it is completed in real time, with borderline or invalid results flagged immediately and corrective action recorded before any patient testing continues.
Standards & compliance context
- This template supports CLIA waived testing documentation by showing that control materials, expiration dates, and corrective actions were checked before results were released.
- It aligns with manufacturer instructions for use, which govern control frequency, acceptable ranges, storage conditions, and open-vial stability for waived assays.
- The record structure also supports broader quality management expectations found in ISO 9001-style document control and internal audit programs.
- If your organization follows FDA Food Code, OSHA, or other local health authority rules for point-of-care workflows, the log can be adapted without losing the core QC traceability.
General regulatory context for orientation only — verify current requirements with counsel or the relevant agency before relying on this template for compliance.
What's inside this template
Session Identification
This section ties the QC run to a specific date, site, device, test, and operator so the session can be traced without ambiguity.
-
QC Log Date
Date on which quality control testing is being performed.
-
Testing Location / Department
Specify the clinic, department, or POC site where testing is performed (e.g., ED, Urgent Care, Lab Draw Station).
-
Instrument / Analyzer Name and ID
Enter the full name and unique instrument ID or serial number of the analyzer being QC’d (e.g., Abbott i-STAT SN-12345, Quidel Sofia 2 SN-67890).
-
Test Name / Analyte
Specify the CLIA-waived test being performed (e.g., Influenza A/B, SARS-CoV-2 Antigen, Urine hCG, Blood Glucose, Group A Strep).
-
Operator Name
Full name of the operator performing QC testing. Must be a trained, authorized user per CLIA 42 CFR §493.1489.
-
Operator Initials
Initials of the operator performing QC testing, used for log authentication.
Reagent and Control Material Verification
This section proves the materials used were in date, properly stored, and appropriate for the run before any control result is accepted.
-
Reagent Kit Lot Number
Record the lot number printed on the reagent kit or test cassette package.
-
Reagent Kit Expiration Date
Record the expiration date on the reagent kit. Do NOT use expired reagents.
-
Reagent kit is within expiration date
Confirm the reagent kit expiration date has not passed as of today’s date. Expired reagents must be removed from service immediately per CLIA 42 CFR §493.1252.
-
Control Material Lot Number
Record the lot number from the QC control material (positive and/or negative control vials/cartridges).
-
Control Material Expiration Date
Record the expiration date on the control material. Include open-vial expiration if applicable per manufacturer instructions.
-
Control material is within expiration date (including open-vial expiry if applicable)
Confirm control material has not expired. Check both printed expiration and any open-vial expiration per manufacturer’s package insert.
-
Reagents and controls stored per manufacturer specifications
Verify storage conditions (temperature, light exposure, humidity) match manufacturer requirements. Check that refrigerated items were not left at room temperature beyond allowed time.
Positive Control Run
This section documents the expected positive response and confirms the assay can detect the target as intended.
-
Positive Control Run Time
Record the time at which the positive control was run.
-
Positive Control Result (Qualitative)
Record the qualitative result of the positive control run (e.g., Positive, Reactive, Detected).
-
Positive Control Result (Quantitative, if applicable)
If the test produces a numeric result, record the quantitative value for the positive control. Leave blank for qualitative-only tests.
-
Positive control result is within manufacturer's acceptable range
Confirm the positive control result meets the expected outcome per the manufacturer’s package insert. A ‘No’ result requires immediate corrective action and suspension of patient testing.
-
Positive Control Photo (if result is borderline or invalid)
Attach a photo of the positive control result strip/display if the result is borderline, invalid, or unexpected.
Negative Control Run
This section documents the expected negative response and helps catch contamination, carryover, or interpretation errors.
-
Negative Control Run Time
Record the time at which the negative control was run.
-
Negative Control Result (Qualitative)
Record the qualitative result of the negative control run (e.g., Negative, Non-Reactive, Not Detected).
-
Negative Control Result (Quantitative, if applicable)
If the test produces a numeric result, record the quantitative value for the negative control. Leave blank for qualitative-only tests.
-
Negative control result is within manufacturer's acceptable range
Confirm the negative control result meets the expected outcome per the manufacturer’s package insert. A ‘No’ result requires immediate corrective action and suspension of patient testing.
-
Negative Control Photo (if result is borderline or invalid)
Attach a photo of the negative control result strip/display if the result is borderline, invalid, or unexpected.
Overall QC Acceptance and Corrective Action
This section records whether the session passed, what was done if it failed, and who was notified before testing continued.
-
Overall QC Status for This Session
Select the overall QC acceptance status based on both positive and negative control results.
-
Corrective Action Taken (if QC failed or invalid)
If QC failed, describe the corrective action taken: e.g., repeated QC with new controls, opened new reagent kit lot, recalibrated instrument, contacted manufacturer technical support, or removed instrument from service. Leave blank if QC was accepted.
-
Repeat QC Performed After Corrective Action
If corrective action was required, was repeat QC performed and accepted before resuming patient testing?
-
Supervisor / Lab Director Notified of QC Failure
If QC failed, confirm that the laboratory supervisor or CLIA laboratory director was notified per facility policy.
-
Additional Comments
Record any additional observations, instrument errors, environmental conditions, or notes relevant to this QC session.
-
Operator Signature
Operator performing QC must sign to attest that all entries are accurate and complete. Required for CLIA record integrity under 42 CFR §493.1105.
How to use this template
- Create one log entry at the start of each QC session and fill in the date, testing location, instrument or analyzer ID, test name, and operator details before running controls.
- Record the reagent kit lot number, expiration date, control material lot number, and open-vial or discard date if applicable, then verify that storage conditions match the manufacturer instructions.
- Run the positive control and negative control exactly as directed by the package insert, and enter the run time plus the qualitative and quantitative results where the assay requires them.
- Compare each control result to the manufacturer’s acceptable range, attach a photo if the result is borderline or invalid, and mark the session failed if any control is out of range.
- If QC fails, document the corrective action taken, repeat the control run after the fix, and notify the supervisor or lab director before patient testing resumes.
- Sign the log after review and file it with your daily QC records so the session can be traced during audits, inspections, or incident review.
Best practices
- Complete the log at the bench while the controls are running so lot numbers, times, and results are captured before memory gaps occur.
- Use the manufacturer’s exact acceptable range and result language for each assay instead of translating it into generic pass or fail terms.
- Flag borderline, faint, or visually ambiguous results with a photo and a note describing the observation at the time of the run.
- Treat open-vial expiration as a separate check from the printed kit expiration date when the package insert requires it.
- Document storage excursions, temperature issues, or transport problems in the comments field even if the control still passes.
- Do not clear a failed QC session until the root cause is addressed and a repeat control run is acceptable.
- Keep the operator signature tied to the person who actually performed the QC, not a later reviewer who only approved the record.
What this template typically catches
Issues teams running this template most often surface in practice:
Common use cases
Frequently asked questions
What does this daily QC log cover?
This template covers one QC session for a CLIA-waived point-of-care test, including session identification, reagent and control verification, positive and negative control results, and final acceptance or corrective action. It is designed to document the exact test, operator, lot numbers, and whether controls were acceptable before patient testing. If your site runs multiple waived assays, you can duplicate the log for each analyte or instrument. It is not a patient result record.
How often should this log be completed?
Use it each day you perform waived testing, and also any time your policy or the manufacturer requires a new QC check, such as after a new lot, new shipment, or instrument restart. Many sites also complete it when controls are borderline, invalid, or when storage conditions may have been compromised. The right cadence is driven by the test manufacturer, your lab policy, and applicable CLIA-waived testing procedures. If a run fails, document the repeat QC after corrective action on the same log.
Who should fill out the log?
The operator who performs the QC run should complete the log and initial the session, since they are the person verifying the control materials and recording the results. A supervisor, lab director, or designated reviewer should be notified if QC fails or if the run cannot be corrected. In small clinics, the same person may perform and review the entry if that is allowed by site policy. The key is that the record clearly shows who ran the QC and who was informed of any failure.
How does this relate to CLIA and other standards?
This log supports CLIA waived testing documentation by showing that controls were run, materials were in date, and corrective action was taken when needed. It also aligns with manufacturer instructions for use, which are the primary source for acceptable control ranges and storage requirements. If your organization follows broader quality systems, the log also fits well with ISO 9001-style traceability and internal audit expectations. It does not replace required competency, training, or maintenance records.
What are the most common mistakes this template helps prevent?
Common problems include recording a control result without the lot number, forgetting to check open-vial expiration, and failing to note whether reagents were stored per manufacturer instructions. Another frequent issue is documenting a failed control without recording the corrective action and repeat run. Borderline or invalid results are also often missed when no photo is attached. This template forces those checks into the workflow so the session is easier to defend during review.
Can I customize this for different waived tests or devices?
Yes. The template is meant to be cloned for each waived assay, such as glucose, urine pregnancy, influenza, strep, or rapid antigen testing, and you can rename fields to match the manufacturer’s terminology. You can also add device-specific prompts for internal quality checks, temperature logs, or cartridge handling if your procedure requires them. Keep the core fields for lot number, expiration, control result, and corrective action so the record stays audit-ready.
Does this replace the manufacturer QC instructions?
No. The log documents that you followed the instructions, but it does not replace them. You still need to use the manufacturer’s acceptable ranges, storage requirements, and run frequency for each test system. If the package insert requires a different control scheme, the log should be adjusted to match it. The safest approach is to mirror the manufacturer workflow inside the template.
How should this be rolled out across a clinic or lab?
Start by assigning one owner for each waived test area, then standardize the fields and acceptable response options before launch. Train operators to complete the log at the time of testing, not after the shift, so lot numbers and control results are captured accurately. Review the first few logs for missing signatures, missing photos on borderline results, and incomplete corrective actions. After that, fold the log into your daily opening checklist or QC routine.
Related templates
Go deeper on the topic
-
A daily huddle is a brief (10–15 minute) standing meeting held at the start of a shift or workday to align the team on priorities, surface issues, and...
-
A deskless worker is any employee whose job happens without a desk, a company laptop, or a fixed workstation. They're roughly 80% of the global workforce —...
-
A frontline employee app is a phone-first application that gives hourly, field, and deskless workers access to their schedule, pay, announcements, training,...
-
A frontline worker is any employee whose job happens away from a desk — on a production floor, in a patient room, behind a store counter, in a customer's...
-
Boost team collaboration with modern tools that improve visibility, accountability, and communication for stronger project outcomes.
-
Discover 4 proven keys to successful project management and team collaboration — from transparent goal-setting to real-time communication and workflow...
-
Compare 9 top shift scheduling platforms for 2026—features, pricing, and workforce fit for frontline, retail, healthcare, and enterprise teams.
-
Intranet file naming conventions that improve search, reduce clutter, and help employees find the right document fast.
Ready to use this template?
Get started with MangoApps and use CLIA Waived Testing Daily Quality Control Log with your team — pricing built for small business.