BAS Sensor Calibration Field Verification
BAS Sensor Calibration Field Verification
Field verification template for comparing BAS temperature, pressure, humidity, and CO2 sensor readings against a calibrated reference instrument and documenting controller offset adjustments, deficiencies, and corrective actions.
Inspection Details and Scope
-
Facility, building, and system identified
Enter the site name, building, BAS system, and controlled area or zone being verified.
-
Date, time, and inspector recorded
Document when the field verification was performed and who completed it.
-
Sensor point and controller point identified
Identify the sensor tag, point name, controller address, and associated BAS loop or zone.
-
Verification scope includes temperature, pressure, humidity, and CO2 as applicable
Select all sensor types included in this verification.
-
Sensor is accessible and safe to test
Confirm the sensor, test port, or sampling location is reachable without unsafe access, exposed conductors, or other hazards.
Reference Instrument Verification
-
Reference instrument make, model, and serial number recorded
Document the calibrated reference instrument used for comparison.
-
Reference instrument calibration is current
Confirm the reference instrument calibration certificate or sticker is valid and not expired.
-
Reference instrument calibration due date recorded
Enter the calibration expiration or due date for the reference instrument.
-
Reference instrument zero or ambient check completed
Verify the reference instrument was checked for zero, ambient, or baseline stability before use.
-
Environmental conditions recorded during verification
Document ambient conditions that may affect readings, such as temperature, airflow, or condensation.
Sensor Readings and Offset Calculations
-
Temperature sensor reading and reference comparison recorded
Enter the BAS temperature reading and document the reference value and calculated offset in the comment field.
-
Pressure sensor reading and reference comparison recorded
Enter the BAS pressure reading and document the reference value and calculated offset in the comment field.
-
Humidity sensor reading and reference comparison recorded
Enter the BAS humidity reading and document the reference value and calculated offset in the comment field.
-
CO2 sensor reading and reference comparison recorded
Enter the BAS CO2 reading and document the reference value and calculated offset in the comment field.
Controller Adjustment and Functional Check
-
Controller offset adjustment recorded
Document the offset value applied at the controller for each adjusted point.
-
Adjustment made at controller or BAS software
Confirm the correction was applied at the controller, point object, or BAS workstation as required by site procedure.
-
Post-adjustment reading within acceptable tolerance
Verify the sensor reading after adjustment is within the site's acceptable tolerance or commissioning criteria.
Deficiencies, Corrective Actions, and Sign-Off
-
Deficiencies or non-conformances documented
Describe any sensor drift, failed comparison, damaged probe, unstable reading, or other deficiency.
-
Corrective action required
Indicate whether follow-up repair, replacement, reprogramming, or re-verification is needed.
-
Inspector signature
Inspector or technician signs to confirm the field verification record is complete and accurate.
Ask AI
Template Studio