Inspection Details

  • Inspection date and time
  • Device identifier / asset tag
  • Inspection location
  • Inspector name and signature

AED Self-Test and Readiness

  • Device self-test passed
    Confirm the AED internal self-test indicates normal operation with no fault or service alert.
  • Readiness indicator shows ready status
    Verify the status indicator is green/ready or otherwise indicates the unit is deployment-ready.
  • No error codes, alarms, or service lights present
    Check for any audible alarms, flashing lights, or displayed error codes requiring service.
  • Cabinet or storage case accessible and unobstructed
    Confirm the AED is mounted or stored in a location that is clearly accessible and not blocked.

Pads, Battery, and Consumables

  • Electrode pads present and sealed
    Verify adult pads are present, unopened, and stored according to manufacturer instructions.
  • Electrode pads expiration date valid
    Record the pad expiration date and confirm it has not passed.
  • Battery installed and within service life
    Confirm the battery is installed, secure, and not due for replacement based on the manufacturer service life or indicator.
  • Battery status indicates sufficient charge
    Verify the battery indicator shows adequate charge for immediate use.
  • Accessory kit present and complete
    Check for required accessories such as scissors, gloves, razor, towel, or barrier device if applicable to the site SOP.

Physical Condition and Documentation

  • Device housing free of visible damage
    Inspect the AED casing, screen, connectors, and cables for cracks, corrosion, missing parts, or other visible damage.
  • Pads and battery replacement dates recorded
    Confirm replacement or expiration dates are documented for tracking and preventive maintenance.
  • Inspection log completed and filed
    Verify this monthly check is documented and retained per site procedure.
  • Corrective action documented for any deficiency
    Describe any non-conformance, removed-from-service condition, or follow-up action taken.
Ask AI Template Studio

Let's customize Defibrillator Monthly Functional Check Log.

Tell me how you'd like to adapt it. For example:

  • Add a question about delivery time.
  • Make it shorter — 5 questions max.
  • Tailor it for the hospitality industry.
  • Translate the labels into Spanish.
Ask AI Product Advisor

Hi! I'm the MangoApps Product Advisor. I can help you with:

  • Understanding our 40+ workplace apps
  • Finding the right solution for your needs
  • Answering questions about pricing and features
  • Pointing you to free tools you can try right now

What would you like to know?