Run: Battery Test Bay Conductance Tester Calibration Audit

Daily audit for a battery test bay conductance tester to confirm the unit is clean, accurate, in calibration, and safe to use before customer or shop testing...

Fill this out, get a PDF emailed to you. No account required. Want to run it with your team and track results? Sign up free →

Tester Identification and Readiness

Verify the tester is located in the designated battery test area and the asset tag/serial number matches the store's assigned equipment record.
Turn the unit on and confirm it boots normally, displays no fault codes, and completes any built-in self-check.
Inspect for cracked housing, unreadable display, damaged buttons, frayed leads, loose connectors, or corroded clamps.
Check the unit and leads for contamination, moisture, or visible electrolyte residue that could affect safe operation or readings.
Confirm the bay area allows safe access to the tester, battery staging area, and customer interaction space without trip hazards or blocked movement.

Calibration and Verification Status

Confirm the unit has a visible calibration/verification label or sticker showing the current status and due date.
Confirm the tester is within the store-required calibration or verification interval and not overdue.
Document that the tester was checked against the approved daily verification method or known-good reference battery per SOP.
If your process uses a reference battery or standard, confirm it is identified, available, and not damaged or out of service.
Enter the date of the most recent annual calibration, health check, or manufacturer service event if required by the store program.

Test Accuracy and Functional Check

Run a sample test or built-in check and verify the screen clearly displays battery condition, voltage, and any required pass/fail or code output.
Record the measured voltage from the verification check. Use the site-approved known-good battery or reference method.
Confirm buttons, touchscreen, or menu navigation respond correctly and do not freeze, lag, or misread inputs.
Verify the clamps grip firmly and the tester does not lose contact or produce intermittent readings during the check.
Compare the displayed result to the expected range or status for the approved verification battery or standard.

Documentation and Records

Confirm the daily calibration audit or verification log was filled out for this unit and date.
Verify the current calibration certificate, service record, or maintenance history is available in the store file or system.
If a problem was found, document the deficiency, assigned owner, and expected completion date.
If the tester fails verification, confirm it is removed from service and clearly tagged to prevent use.

Closeout and Sign-Off

Confirm escalation occurred for any critical item failure, overdue calibration, or out-of-service condition.
Add any notes about unusual readings, maintenance needs, or follow-up actions.
Signature of the person completing the daily audit.

Get your results

Enter your email — we'll send you a PDF of your filled-out template, plus the occasional MangoScoop newsletter (templates, workflow tips, product updates). Unsubscribe anytime — link is in every email.

Generated with MangoApps Templates — browse 250+ free
Ask AI Product Advisor

Hi! I'm the MangoApps Product Advisor. I can help you with:

  • Understanding our 40+ workplace apps
  • Finding the right solution for your needs
  • Answering questions about pricing and features
  • Pointing you to free tools you can try right now

What would you like to know?