Anesthesia Machine Pre-Use Check
Anesthesia Machine Pre-Use Check
Pre-use inspection template for anesthesia machines covering the high-pressure system, low-pressure system, breathing circuit, and gas analyzer checks.
Machine Identification and Pre-Use Setup
-
Anesthesia machine identified and assigned to the correct procedure room
Confirm the machine is the intended unit for the case and is positioned for immediate use.
-
Power connected and machine passes startup self-test
Verify the unit powers on normally and any built-in self-check completes without unresolved alarms or errors.
-
Vaporizers secured and agent level within usable range
Confirm vaporizers are locked in place, correctly filled if applicable, and set according to local policy.
High-Pressure System
-
Oxygen supply cylinder present, secured, and full enough for emergency use
Check cylinder presence, secure mounting, and pressure adequate for backup oxygen supply.
-
Cylinder pressure gauge indicates acceptable pressure
Record the cylinder pressure reading and confirm it meets local minimum backup requirements.
-
Pipeline gas hoses connected to correct wall outlets and secure
Verify oxygen, nitrous oxide, and air hoses are connected to the correct outlets and are not loose, damaged, or crossed.
-
Pipeline pressure indicators within normal operating range
Confirm pipeline supply pressures are normal and no low-supply alarms are present.
Low-Pressure System
-
Flow control knobs move smoothly and return to closed position
Check each flow control knob for smooth operation and proper closure.
-
Flowmeters are intact, readable, and function during flow adjustment
Verify flowmeters are not cracked, stuck, or obscured and that bobbins respond appropriately.
-
Low-pressure system leak test completed without leak
Perform the machine leak test per local SOP and confirm the system holds pressure.
-
Oxygen flush valve operates and returns normally
Confirm the oxygen flush delivers flow when activated and stops when released.
Breathing Circuit and Ventilator
-
Breathing circuit assembled correctly and free of visible damage
Inspect hoses, connectors, Y-piece, and reservoir bag for cracks, disconnections, or wear.
-
Reservoir bag inflates and deflates appropriately
Confirm the bag responds normally during manual ventilation and system testing.
-
APL valve functions and is set appropriately
Verify the adjustable pressure-limiting valve turns freely and is set per case requirements.
-
CO2 absorbent present and not exhausted or channeling
Inspect absorbent canister condition, color change, and evidence of channeling or depletion.
-
Ventilator performs test cycle without alarm or abnormal pressure
Run the ventilator test mode and verify normal cycling, pressures, and alarm behavior.
Gas Analyzer and Monitoring
-
Gas analyzer powers on and completes calibration or warm-up
Confirm the analyzer is operational and ready to display accurate readings.
-
Inspired and end-tidal gas readings display correctly
Verify the monitor displays oxygen, anesthetic agent, and carbon dioxide values as applicable.
-
Gas sampling line connected and unobstructed
Check the sampling line for secure connection, kinks, moisture, or occlusion.
-
Alarm limits set according to local policy
Confirm high and low alarm thresholds are configured before patient use.
Ask AI
Template Studio